TABLE 1.
Parameter | Value for IgG detection in study 1 |
|||||||
---|---|---|---|---|---|---|---|---|
Nucleocapsid-coated plates |
Spike (S1)-coated plates |
RBD-coated plates |
cPass sVNT |
|||||
Positive (n = 45) | Negative (n = 23) | Positive (n = 45) | Negative (n = 23) | Positive (n = 45) | Negative (n = 23) | Positive (n = 45) | Negative(n = 23) | |
No. of positive samples | 28 | 1 | 34 | 0 | 43 | 3 | 39 | 0 |
No. of negative | 17 | 22 | 11 | 23 | 2 | 20 | 6 | 23 |
Sensitivity (%) | 62.22 | 75.56 | 95.56 | 86.67 | ||||
Specificity (%) | 95.65 | 100.00 | 86.96 | 100.00 | ||||
Accuracy (%) | 92.31 | 97.36 | 87.82 | 98.67 | ||||
PPV (%) | 61.39 | 100.00 | 44.87 | 100.00 | ||||
NPV (%) | 95.80 | 97.36 | 99.44 | 98.54 |
See Fig. 2. Sensitivity, specificity, accuracy, positive predictive values (PPV), and negative predictive values (NPV) are shown. A prevalence of 10% was used for the calculations.