. 2021 Mar 19;59(4):e02438-20. doi: 10.1128/JCM.02438-20

TABLE 1.

Combined data from study 1 comparing assay performances of three commercial serology tests and the cPass sVNTa

Parameter Value for IgG detection in study 1
Nucleocapsid-coated plates
Spike (S1)-coated plates
RBD-coated plates
cPass sVNT
Positive (n = 45) Negative (n = 23) Positive (n = 45) Negative (n = 23) Positive (n = 45) Negative (n = 23) Positive (n = 45) Negative(n = 23)
No. of positive samples 28 1 34 0 43 3 39 0
No. of negative 17 22 11 23 2 20 6 23
Sensitivity (%) 62.22 75.56 95.56 86.67
Specificity (%) 95.65 100.00 86.96 100.00
Accuracy (%) 92.31 97.36 87.82 98.67
PPV (%) 61.39 100.00 44.87 100.00
NPV (%) 95.80 97.36 99.44 98.54
a

See Fig. 2. Sensitivity, specificity, accuracy, positive predictive values (PPV), and negative predictive values (NPV) are shown. A prevalence of 10% was used for the calculations.