. 2021 Mar 19;59(4):e02438-20. doi: 10.1128/JCM.02438-20

TABLE 2.

Combined data from study 2 comparing the assay performances of six commercial serology tests and the cPass sVNTa

Assay Analyte Time after symptom onset
Negative samples (serum collected before Nov 2019)
PPV (%) NPV (%) ACC (%)
5–9 days
10–19 days
>19 days
All time points
No. of Neg samples No. of Equ samples No. of Pos samples Spec (%) CI (%)
No. of Neg samples No. of Pos samples Sens (%) CI (%) No. of Neg samples No. of Pos samples Sens (%) CI (%) No. of Neg samples No. of Pos samples Sens (%) CI (%) No. of Neg samples No. of Pos samples Sens (%) CI (%) Total no. of samples
Abbott IgG 0 6 100 54–100 3 9 75 43–95 0 9 100 66–100 3 24 89 71–98 27 49 0 1 98 89–100 83 99 97
Affinity IgG 0 7 100 59–100 2 11 85 55–98 0 8 100 63–100 2 26 93 71–98 28 47 0 0 100 92–100 100 99 99
Bio-Rad IgG 0 6 100 54–100 3 9 75 43–95 0 9 100 66–100 3 24 89 71–98 27 50 0 0 100 93–100 100 99 99
DiaSorin IgG 4 2 33 4–78 4 8 67 35–90 0 9 100 66–100 8 19 70 71–98 27 48 1 1 96 86–100 66 97 93
Euroimmun IgG 3 3 50 12–88 4 8 67 35–90 0 8 100 63–100 7 19 73 71–98 26 50 0 0 100 93–100 100 97 97
Roche Total Ab 1 6 86 42–100 4 9 69 39–91 0 9 100 66–100 5 24 83 71–98 29 50 0 0 100 93–100 100 98 98
GenScript RBD-hACE2 1 6 86 42–100 1 12 92 64–100 0 9 100 66–100 2 27 93 71–98 29 50 0 0 100 93–100 100 99 99
a

Sensitivity (Sens), specificity (Spec), accuracy (ACC), positive predictive values (PPV), and negative predictive values (NPV) are shown. Pos, positive; Neg, negative; CI, 95% confidence interval; Equ, equivocal result; Ab antibody. A prevalence of 10% was used for the calculations.