Table 2.
95% Limits of agreement (difference) between the semi-automatic methods and the ROs for both types of treatment.
| Left - Right (cm) | Superior - Inferior (cm) | Anterior - Posterior (cm) | |||||
|---|---|---|---|---|---|---|---|
| Prostate irradiation | CM + MM | −0.15 | +0.22 | −0.33 | +0.23 | −0.32 | +0.23 |
| DRT | −0.18 | +0.23 | −0.47 | +0.25 | −0.52 | +0.24 | |
| Prostate bed irradiation | CM + MM | −0.16 | +0.22 | −0.23 | +0.16 | −0.23 | +0.16 |
| DRT | −0.15 | +0.23 | −0.36 | +0.23 | −0.38 | +0.31 | |
DRT: Dual registration tool; CM + MM: Chamfer match + manual match.
Figures in bold exceeded predefined threshold.