Skip to main content
. 2020 Jan 6;97(2):199–201. doi: 10.1002/cyto.a.23962

Table 2.

Corrected instrument configuration corresponding to online Table 1

Laser Wavelength (nm) Laser Power (mW) Laser Type Detector Spectral Range for Detector (nm) Dichroic LP Mirror (nm) Band Pass (nm) Example Fluorochrome
637 (Red) 200 DPSS R780 750–810 750 780/60 APC‐Cy7
R710 723–753 685 730/45 Alexa700
R660 655–685 630 670/30 APC
532 (Green) 200 DPSS G780 750–810 750 780/60 PE‐Cy7
G710 685–735 690 710/50 PE‐Cy5.5
G660 640–680 635 660/40 PE‐Cy5
G610 600–620 600 610/20 PE‐Dazzle594
G575 563–588 None 575/25 PE
488 (Blue) 200 DPSS B780 760–800 740 780/40 BB790
B710 685–735 690 710/50 BB700
B660 640–680 635 660/40 BB660
B610 600–620 600 610/20 BB630
B515 505–525 505 515/20 FITC
405 (Violet) 200 DPSS V780 754–816 770 785/62 BV785
V750 735–765 735 750/30 BV750
V710 690–730 685 710/40 BV711
V655 650–670 630 660/20 BV650
V682.510110 595–615 580 605/20 BV605
V570 563–588 550 575/25 BV570
V510 505–525 505 515/20 BV510
V450 430–470 None 450/40 V450
355 (Ultraviolet) 60 DPSS U780 754–816 755 785/62 BUV805
U730 723–758 710 740/35 BUV737
U660 640–680 635 660/40 BUV661
U570 550–590 550 570/40 BUV563
U500 500–530 470 515/30 BUV495
U450 425–475 410 450/50 Viability
U395 365–393 None 379/28 BUV395