Skip to main content
. 2021 Apr 20;12:645085. doi: 10.3389/fendo.2021.645085

Table 1.

ELISA performance characteristics.

Vasoinhibin standard curve in all sandwiches 0.1 – 20 µg/ml
Applied LOQ (lowest calibration standard) 100 ng
Intra-assay coefficient of variation of standard 12.5%
Inter-assay coefficient of variation of standard 14%
Cross-reactivity with PRL at 12 µg/ml at 5.5 µg/ml vasoinhibin standard
(% deviation compared to vasoinhibin standard only)
C50-C10 = 3.3%
C10-C50 = 7.3%
Sandwich BLK [O.D.] (mean ± SD) LOD [O.D.] LOD - BLK [O.D.] LOQ [O.D.] LOQ – BLK [O.D.]
C1F-C50H 0.096 ± 0.011 0.129 0.033 0.206 0.110
C10F-C50H 0.768 ± 0.036 0.876 0.108 1.128 0.360
C50H-C10F 0.725 ± 0.071 0.938 0.213 1.435 0.710

BLK, blank; LOD, limit of detection; BLK mean plus the three-fold standard deviation of blank, LOQ, limit of quantitation, the applied limit of quantitation was the lowest calibration standard of 100 ng, the LOQ is the BLK-mean plus the 10-fold standard deviation of blank. LOD and LOQ are also shown as BLK-subtracted values for comparison with the standard curves.