Table 1.
Vasoinhibin standard curve in all sandwiches | 0.1 – 20 µg/ml | |||||
---|---|---|---|---|---|---|
Applied LOQ (lowest calibration standard) | 100 ng | |||||
Intra-assay coefficient of variation of standard | 12.5% | |||||
Inter-assay coefficient of variation of standard | 14% | |||||
Cross-reactivity with PRL at 12 µg/ml at 5.5 µg/ml vasoinhibin standard (% deviation compared to vasoinhibin standard only) |
C50-C10 = 3.3% C10-C50 = 7.3% |
|||||
Sandwich | BLK [O.D.] (mean ± SD) | LOD [O.D.] | LOD - BLK [O.D.] | LOQ [O.D.] | LOQ – BLK [O.D.] | |
C1F-C50H | 0.096 ± 0.011 | 0.129 | 0.033 | 0.206 | 0.110 | |
C10F-C50H | 0.768 ± 0.036 | 0.876 | 0.108 | 1.128 | 0.360 | |
C50H-C10F | 0.725 ± 0.071 | 0.938 | 0.213 | 1.435 | 0.710 |
BLK, blank; LOD, limit of detection; BLK mean plus the three-fold standard deviation of blank, LOQ, limit of quantitation, the applied limit of quantitation was the lowest calibration standard of 100 ng, the LOQ is the BLK-mean plus the 10-fold standard deviation of blank. LOD and LOQ are also shown as BLK-subtracted values for comparison with the standard curves.