Equipment use and decontamination recommendations.

Equipment Use Decontamination
Flow sensors Reusable (discard after 10 uses) Wash with soap and water. Submerge in neutral-pH enzymatic solution. Sterile processing recommended.
Mouthpiece Reusable Wash with soap and water. Soak for 60 min in 10% bleach solution and rinse with cold water; OR Submerge in neutral-pH enzymatic solution and send for sterile processing.
Nose clip Single use Discard after use; OR wash with soap and water. Soak for 60 min in 10% bleach solution and rinse with cold water.
Face mask and coupler Reusable Wash with soap and water. Soak for 10 min in 10% bleach solution and rinse with cold water; OR Submerge in neutral-pH enzymatic solution and send for sterile processing.
Neoprene masks Single patient usea Wash with warm soap and water. Rinse with cold water. Air dry.
Pulse oximetry band/ head band Single patient usea Wash with warm soap and water. Rinse with cold water. Air dry.
a

Can be reused for same patient and discard when tests are discontinued.