4 |
Case cover will not slide off |
Not all screws have been removed |
Check both the back and the bottom of the
projector to ensure all necessary screws have been removed |
9 |
Filters will not fit |
Mis-sized filter |
Measure the opening at the location for the
filters and check dimension of the custom filters. Alter the filters as
necessary to fit. |
14 |
Projector will not turn on |
Internal disconnected cables |
If all cables were not correctly connected
when reassembling projector, then unit will not power up. Take projector
apart and ensure all cables are connected. |
|
|
Other errors associated with projector |
Check error blinking codes and consult
user’s manual for solution |
|
Dim image |
Bulb is near the end of life |
Check projector for bulb hours and consult
user’s manual on replacing |
|
Color is missing/absent |
Disconnected LCD panel cable |
Disassemble projector to ensure that LCD panel
cables have been securely reattached |
|
|
Shifted or broken filter |
Disassemble projector and check all inserted
filters to check if they have shifted or possibly broken. Replace if
necessary. |
|
Color or image is striped |
Unsecured/loose LCD panel cable |
Disassemble projector to ensure that LCD panel
cables have been securely reattached |
18.A.v |
Insufficient space to position the
projector |
Accessory tube lens focal plane located within
body of microscope |
Extend the accessory tube lens to the rear of
the microscope allowing the focal plane to be located outside the body
of the microscope |
18.B.v |
Insufficient space to position the
projector |
Relay lens focal plane located within body of
microscope |
Select a matched lens pair of greater focal
length |
21 |
Image never focuses |
Mis-positioned lenses |
Check that all lenses are located as described
in Figure 1. |
24.C.i |
Program gives error upon start |
Altered camera from what program was written
for |
Alter the LabVIEW code to communicate with the
specific camera used. See also supporting documentation of Ref. 12. |
24.C.ix |
Program gives error upon start |
Altered camera from what program was written
for |
Alter the LabVIEW code to communicate with the
specific camera used. See also supporting documentation of Ref. 12. |
|
|
Altered motorized stage from what program was
written for |
Alter the LabVIEW code to communicate with the
specific stage used. See also supporting documentation of Ref. 12. |