Skip to main content
. 2021 Apr 14;413(13):3561–3571. doi: 10.1007/s00216-021-03307-6

Table 2.

Summary of validation parameters obtained from clavulanic acid calibration curves in ACN, plasma, and meat

ACN Plasma Meat
Linearity
  Concentration range (μg·L−1/μg·kg−1) 10–2000 10–2000 50–2000
  Number of points 6 6 7
  Correlation coefficient, R2 0.9929 0.9991 0.9918
  Standard deviation of the correlation coefficient 0.2857 0.1998 0.2217
  Slope, m 0.0064 0.0113 0.0036
  Standard deviation of the slope, Sm 0.00059 0.00034 0.00008
  Intercept, b −0.0089 −0.1093 0.0060
  Standard deviation of the intercept, Sb 0.0974 0.0587 0.0688
  LOD (μg·L−1/μg·kg−1) 3.18 3.09 2.57
  LOQ (μg·L−1/μg·kg−1) 10.48 10.21 8.47
Accuracy
  Recovery (%) 94.6 105.7 95.6
Within-run precision
  Concentration level (μg·L−1/μg·kg−1) 10 500 2000 10 500 2000 50 400 2000
  Calculated concentration (μg·L−1/μg·kg−1) 7.71 513.73 2076.28 12.3 518.96 2088.70 45.58 399.18 1955.81
  CV (%) 13.6 6.5 7.7 10.9 2.8 5.3 8.5 6.5 6.9