Table 1. Recommendations for 99mTc-pyrophosphate scintigraphy acquisition parameters.

Parameters
Imaging procedure
 Specific preparation None; fasting is not required
 Scan Rest scan
 Dose of 99mTc-PYP • Recommended: 20 mCi for 3-hour acquisition
• Adjust according to different models of instrument
 Time interval between injection and acquisition • Recommended: 3-hour planar and SPECT
• Optional: 1-hour planar
Imaging parameters
 Field of view Cardiac or chest
 Image type Cardiac or chest planar and SPECT (or SPECT/CT)
 Position Supine
 Energy window 140 keV, 15-20%
 Collimators Low energy, high resolution
 Matrix 128 × 128 recommended (64 × 64 minimum required) for SPECT, 256 × 256 for planar
 Pixel size 2.3-6.5 mm (may change with different matrix)
Planar imaging specific parameters
 Number of views Anterior, lateral, and left anterior oblique (LAO)
 Detector configuration 180 degrees
 Image duration 750,000 counts
 Magnification 1.45-1.5
SPECT imaging specific parameters
 Angular range 360 degrees
 Detector configuration 180 degrees
 ECG gating Off; non-gated imaging
 Number of views/detectors 32-64
 Time per stop 20 seconds

ECG, electrocardiography; keV, kilo-electronvolts; mCi, millicurie; PYP, pyrophosphate; SPECT, single-photon emission computed tomography; SPECT/CT, SPECT/computed tomography.