Table 1. Recommendations for 99mTc-pyrophosphate scintigraphy acquisition parameters.
Parameters | |
Imaging procedure | |
Specific preparation | None; fasting is not required |
Scan | Rest scan |
Dose of 99mTc-PYP | • Recommended: 20 mCi for 3-hour acquisition |
• Adjust according to different models of instrument | |
Time interval between injection and acquisition | • Recommended: 3-hour planar and SPECT |
• Optional: 1-hour planar | |
Imaging parameters | |
Field of view | Cardiac or chest |
Image type | Cardiac or chest planar and SPECT (or SPECT/CT) |
Position | Supine |
Energy window | 140 keV, 15-20% |
Collimators | Low energy, high resolution |
Matrix | 128 × 128 recommended (64 × 64 minimum required) for SPECT, 256 × 256 for planar |
Pixel size | 2.3-6.5 mm (may change with different matrix) |
Planar imaging specific parameters | |
Number of views | Anterior, lateral, and left anterior oblique (LAO) |
Detector configuration | 180 degrees |
Image duration | 750,000 counts |
Magnification | 1.45-1.5 |
SPECT imaging specific parameters | |
Angular range | 360 degrees |
Detector configuration | 180 degrees |
ECG gating | Off; non-gated imaging |
Number of views/detectors | 32-64 |
Time per stop | 20 seconds |
ECG, electrocardiography; keV, kilo-electronvolts; mCi, millicurie; PYP, pyrophosphate; SPECT, single-photon emission computed tomography; SPECT/CT, SPECT/computed tomography.