Table 2. Differences in 99mTc-pyrophosphate scintigraphy acquisition parameters between ASNC* and TSOC/SNMROC consensus recommendations.

TSOC and SNMROC recommendations 2019 ASNC practice points recommendations*
Imaging procedure
 Dose of 99mTc-PYP • Recommended: 20 mCi for 3-hour acquisition • 10-20 mCi
• Adjust according to different models of instrument
 Time interval between injection and acquisition • Recommended: 3-hour planar and SPECT • Recommended: 1-hour planar and SPECT
• Optional: 1-hour planar • Optional: 3-hour SPECT or planar
Imaging parameters
 Matrix • Recommended: 128 × 128 for SPECT 64 × 64 minimum
• Required: 64 × 64 minimum for SPECT
• 256 × 256 for planar
Planar imaging specific parameters
 Detector configuration 180 degrees 90 degrees
 Magnification 1.45-1.5 1.46
SPECT imaging specific parameters
 Number of views/detector 32-64 40

ASNC, American Society of Nuclear Cardiology; ECG, electrocardiography; mCi, millicurie; PYP, pyrophosphate; SNMROC, Society of Nuclear Medicine of the Republic of China; SPECT, single-photon emission computed tomography; TSOC, Taiwan Society of Cardiology.

* Source: Dorbala S, et al. 2019 ASNC Practice Points.26