Figure 6.

Figure 6.

(A) Apparent mobility vs electric field strength of rNMPs in 110 nm x 110 nm nanochannels fabricated in COC/COC nanochannels using 1X NEB buffer 3 at pH 7.9 as the carrier electrolyte. (B) Histogram of apparent mobilities of rNMPs at 930 V/cm in 110 × 110 nm COC/COC nanochannel devices using 1X NEB buffer 3 at pH 7.9 as the carrier electrolyte. The histograms were fit to Gaussian functions and each bin width represented 2 × 10−6 cm2/Vs. (C) The variance (σ2) of peaks estimated from the Gaussian fits to the histogram data shown in (B). (D) The resolution of Gaussian fits calculated using R = 1.18 (Δμ/w0.5 + w0.5), where w0.5 correspond to the full width at half maximum of the Gaussian fits. (E) Identification accuracies of rNMPs calculated from Gaussian fit overlap.