Skip to main content
. 2021 May 10;294:114182. doi: 10.1016/j.jviromet.2021.114182

Table 2.

Determination of the 95 % limit of detection (LoD) of the Aptima SARS-CoV-2 assay.

SARS-CoV-2 Copies / ml Positive/total number tested (n) Positive rate (%)
1×104 9/9 100 %
1×103 9/9 100 %
1×102 17/27 63.0%
1×101 1/27 3.7%
1×10° 0/27 0%