. 2012 Mar 5;14(4):222–227. doi: 10.1111/j.1751-7176.2012.00599.x

Table I.

 Descriptive Statistics for Measurement Device, Setting, Patient Factor Observations, and Techniques Influencing Measurement Accuracy (N=119)

No. (%)
Measurement device
 Dinamap 84 (71)
 Mercury 12 (10)
 Phillips 9 (7.5)
 Welch‐Allyn 9 (7.5)
 Dinamap‐antiquated 5 (4)
Measurement setting
 Central 45 (38)
 Private 68 (57)
 Semi‐private 6 (5)
Room temperature
 Cool 4 (3)
 Ambient 104 (88)
 Warm 11 (9)
Observations
 Measurement of arm circumference 0 (0)
 Clothing, sleeve, or garment interference 30 (25)
 Out of position/bad posture 18 (15)
 Improper device use 7 (6)
 Stethoscope bell use (where appropriate, *n=12) 9a (8)
 Conversation (yes) 48 (40)
 Bare arm (no) 49 (41)
 Arm/cuff at heart level (no) 89 (75)
 Feet not flat (no) 77 (65)
 Crossed legs (yes) 24 (20)
 Second blood pressure measurement (no) 114 (95)
 Clinic rest time (no) 117 (98)