Method | Interassay Coefficients of Variation | Detection Limit for Albumin | False‐Negative vs HPLC, % | False‐Positive vs HPLC, % |
---|---|---|---|---|
HPLC | 2.4% at 95.8 mg/L | 2 mg/L | ND | ND |
Immunonephelometry (Beckman Array Analyzer; Global Medical Instrumentation, Inc, Ramsey, MN) | 4.2% at 12.1 mg/L and 5.3% at 45 mg/L | 2 mg/L | ND | ND |
Immunoturbidity (Dade‐Behring Turbimeter; Dade Behring, Inc, Deerfield, IL) | 4.1% at 10.6 mg/L and 2.2% at 77.9 mg/L | 6 mg/L | 36 | 0 |
Radioimmunoassay | 9.2% at 12.2 mg/dL | 16 µg/L | 23 | 0 |