Skip to main content
. 2021 Jan 21;59(2):e02257-20. doi: 10.1128/JCM.02257-20

TABLE 3.

Concordance between manufacturer enzyme immunoassay cutoff values for SARS-CoV-2 seropositivity and high nAb titers at various thresholds (n = 140)a

Serologic assay Positive percent agreement, no. (%)
Negative percent agreement, no. (%)
nAb ≥20
(n = 92)
nAb ≥40
(n = 80)
nAb ≥80
(n = 64)
nAb ≥160
(n = 35)
nAb <20
(n = 48)
nAb <40
(n = 60)
nAb <80
(n = 76)
nAb <160
(n = 105)
Euroimmun 90 (97.8) 80 (100) 64 (100) 35 (100) 16 (33.3) 18 (30.0) 18 (23.7) 18 (17.1)
EDI 86 (93.5) 74 (92.5) 61 (95.3) 34 (97.1) 25 (52.1) 25 (41.7) 28 (36.8) 30 (28.6)
ImmunoDiagnostics 86 (93.5) 76 (95.0) 61 (95.3) 35 (100) 27 (56.3) 29 (48.3) 30 (39.5) 33 (31.4)
Abbott 90 (97.8) 79 (98.8) 64 (100) 35 (100) 9 (18.8) 10 (16.7) 11 (14.5) 11 (10.5)
Roche 90 (98.4) 78 (97.5) 63 (98.4) 34 (97.4) 6 (12.5) 6 (10.0) 7 (9.2) 7 (6.7)
a

Four thresholds for a high neutralizing antibody (nAb) AUC value were examined as the reference positive test. Borderline/indeterminate specimens were considered negative for the EIAs per manufacturers’ cutoff values.