Table 1.

Abbott Alinity m Clinical Concordance with Roche cobas 6800 and Hologic Panther Fusion SARS-CoV-2 Assay.

Roche cobas 6800 Hologic Panther Fusion
Positive Negative Total Positive Negative Total
Positive 83 0 83 30 0 30 Abbott Alinity m #1
Negative 0 106 106 0 90 90
Total 83 106 189 30 90 120
Positive Negative Total Positive Negative Total
Positive 94 1 95 30 0 30 Abbott Alinity m #2
Negative 0 89 89 0 84 84
Total 94 90 184 30 84 114

The positive percent agreement (PPA) and for both Alinity m instruments was reported at 100% compared to the cobas 6800 SARS-CoV-2 assay. The negative percent agreement (NPA) was determine at 100% and 98.9% for instrument #1 and #2, respectively. In comparison to the Panther Fusion SARS-CoV-2 assay, both PPA and NPA were calculated at 100%. Line-item CN values for all samples on each platform are included in Supplemental Data 1.