Table 1:
Variable | Aquilion 64 | Brilliance 64 | LightSpeed 64 | Sensation 64 |
---|---|---|---|---|
Acquisition protocol | ||||
Tube voltage, kV | 120 | 140 | 140 | 120 |
Tube current, mA | 200 | 200 | 335 | 135 |
Beam collimation, mm | 4 × 0.5 | 2 × 0.55 | 32 × 0.625 | 12 × 0.6 |
Pitch | 0.75 | 0.5 | 0.531 | 0.45 |
Rotation time, s | 0.5 | 0.5 | 0.6 | 1.0 |
Scan FOV | 240 | 500 | 320 | 500 |
Dose of acquisition protocol | ||||
Effective dose, mSv | 1.4 | 2.0 | 1.8 | 1.3 |
Reconstruction protocol | ||||
Section thickness, mm | 0.5 | 0.55 | 0.6 | 0.6 |
Section interval, mm | 0.3 | 0.3 | 0.3 | 0.3 |
Kernel | FC84 | Filter D | BonePlus | U90u |
Reconstruction matrix | 5122 | 7682 | 5122 | 5122 |
Recon FOV PMMA phantom, mm | 100 | |||
Recon FOV PSF phantom, mm | 50 | |||
Recon FOV cadaver head, mm | 80 |
Note:—FOV indicates field of view; PMMA, polymethylmethacrylate; PSF, point spread function.