. 2008 Aug;29(7):1382–1388. doi: 10.3174/ajnr.A1108

Table 1:

Acquisition and reconstruction protocols and effective radiation dose for cochlear implant imaging on 64-section scanners

Variable Aquilion 64 Brilliance 64 LightSpeed 64 Sensation 64
Acquisition protocol
    Tube voltage, kV 120 140 140 120
    Tube current, mA 200 200 335 135
    Beam collimation, mm 4 × 0.5 2 × 0.55 32 × 0.625 12 × 0.6
    Pitch 0.75 0.5 0.531 0.45
    Rotation time, s 0.5 0.5 0.6 1.0
    Scan FOV 240 500 320 500
Dose of acquisition protocol
    Effective dose, mSv 1.4 2.0 1.8 1.3
Reconstruction protocol
    Section thickness, mm 0.5 0.55 0.6 0.6
    Section interval, mm 0.3 0.3 0.3 0.3
    Kernel FC84 Filter D BonePlus U90u
    Reconstruction matrix 5122 7682 5122 5122
    Recon FOV PMMA phantom, mm 100
    Recon FOV PSF phantom, mm 50
    Recon FOV cadaver head, mm 80

Note:—FOV indicates field of view; PMMA, polymethylmethacrylate; PSF, point spread function.