Skip to main content
. 2021 Jan 4;50(7):1389–1397. doi: 10.1007/s00256-020-03702-7

Table 3.

Percentage of correct prediction of the true value (3D-CT = gold standard) using Kanekasu view radiographs and 2-dimensional computer tomography (2D-CT) within 2 or 3 degrees of variability. The percentages refer to measurements based on the anatomical transepicondylar axis

Predicting within ± 2° within ± 3°
2D-CT by Kanekasu 66/82 = 80.5% 75/82 = 91.5%
3D-CT by Kanekasu 54/82 = 65.9% 67/82 = 81.7%
3D-CT by 2D-CT 68/82 = 82.9% 75/82 = 91.5%