Table 2:

Summary of volumetric absorptive microsampling (VAMS)-based liquid chromatography with tandem mass spectrometry (LC-MS/MS) assays for determining the concentration of antimicrobial drugs

Drug Assay range (μg/mL) Sample collection and processing Method validation and clinical application
Cefepime23 0.1 – 100 Sample volume: 10 μL
Drying time: 1 h at room temperature (RT)
Extraction solvent: Water (rehydration); acetonitrile
Inter-assay accuracy: 100 – 108%.
Inter-assay precision: 4.27 – 12.2%.
Recovery (at 3 hematocrit (HCT) levels): Quality control (QC) low 40.8 – 60.2%; QC medium 43.4 – 57.8%; QC high 48.2 – 62.1%
Matrix effect: QC low 96.7%; QC medium 89.5%; QC high 89.8%
HCT effect: recovery decreases as HCT increase
Stability: 1 week at 4°C; 1 month at −20°C; 3 months at −78°C
Clinical application: 1 subject; 7 samples
Piperacillin17 3.125 – 200 Sample volume: 10 μL
Drying time: 1 h at RT
Extraction solvent: Water (rehydration); methanol
Inter-assay accuracy: 94.9 – 106.0%.
Inter-assay precision: 3.2 – 13.6%.
Recovery: QC low 86%; QC high 90%
Matrix effect: QC low 109%; QC high 107%
HCT effect: Bias <15% for HCT levels 30 – 60%
Stability: 72 h at RT, 4°C and −20°C; 1 month at −20°C; freeze-thaw acceptable
Clinical application: 1 subject; 2 samples
Tazobactam17 3.125 – 200 Sample volume: 10 μL
Drying time: 1 h at RT
Extraction solvent: Water (rehydration); methanol
Inter-assay accuracy: 92.0 – 115.7%.
Inter-assay precision: 6.2 – 9.1%.
Recovery: QC low 88%; QC high 83%
Matrix effect: QC low 101%; QC high 110%
HCT effect: Bias <15% for HCT levels 30 – 60%
Stability: 72 h at RT, 4 °C and −20 °C; 1 month at −20°C; freeze-thaw acceptable
Clinical application: 1 subject; 2 samples
Meropenem17 3.125 – 200 Sample volume: 10 μL
Drying time: 1 h at RT
Extraction Solvent: Water (rehydration); methanol
Inter-assay accuracy: 93.5 – 105.3%.
Inter-assay precision: 7.1 – 10.6%.
Recovery: QC low 86%; QC high 78%
Matrix effect: QC low 95%; QC high 109%
HCT effect: Bias <15% for HCT levels 30 – 60%
Stability: 72 h at RT, 4°C and −20°C; 1 month at −20°C; freeze-thaw acceptable
Clinical application: 3 subjects; 6 samples
Ceftazidime17 3.125 – 200 Sample volume: 10 μL
Drying time: 1 h at RT
Extraction solvent: Water (rehydration); methanol
Inter-assay accuracy: 95.2 – 115.7%.
Inter-assay precision: 0.6 – 5.3%.
Recovery: QC low 103%; QC high 86%
Matrix effect: QC low 104%; QC high 97%
HCT effect: Bias <15% for HCT levels 30 – 60%
Stability: 72 h at RT, 4°C and −20°C; 1 month at −20°C; freeze-thaw acceptable
Clinical application: 3 subjects; 6 samples
Linezolid17 3.125 – 200 Sample volume: 10 μL
Drying time: 1 h at RT
Extraction solvent: Water (rehydration); methanol
Inter-assay accuracy: 91.6 – 109.3%.
Inter-assay precision: 2.3 – 9.5%.
Recovery: QC low 94%; QC high 83%
Matrix effect: QC low 111%; QC high 95%
HCT effect: Bias <15% for HCT levels 30 – 60%
Stability: 72 h at RT, 4°C and −20°C; 1 month at −20°C; freeze-thaw acceptable
Clinical application: 5 subjects; 10 samples