Table 2.

Comparison of commercially available closed-loop systems using the CARES paradigm (32).

CARES 670G/780G Control IQ CamAPS FX
Licensing 7 years+ 6 years+ 1 year+
Availability USA and Europe USA & Europe Europe
Pump Medtronic 670G / 780G Tandem t:slim X2 Dana RS, Dana-i
Insulin Rapid-acting Rapid-acting Ultra-rapid and rapid-acting
Closed-loop term Auto Mode Control IQ Auto mode
Calculate Algorithm Treat-to-target proportional integral derivative with insulin feedback (670G); added fuzzy logic component (780G) Treat-to-range predictive control Treat-to-target MPC
Set-up TDD, weight, basal rates, ICR, ISF, active insulin time.
7 days of manual mode
TDD, weight, basal rates, ICR, ISF TDD and weight
Adaptive learning Overall Not specified by manufacturer Overall, diurnal, meals
Automated insulin delivery Based on total daily insulin dose last 2-6 days Based on pre-programmed basal rates Based on adaptive learning
Automated corrections used to supplement basal delivery 670G: No780G: Yes Yes No
Adjust Glucose target 670G: Target 6.7mmol/L non-customisable.780G: Customisable target of 5.5, 6.1, or 6.7mmol/L. Target range 6.2 – 8.9mmol/L.Sleep range 6.2 – 6.7mmol/L.Non-customisable. Target 5.8mmol/L.Customisable between 4.4-11mmol/L,adjustable in 0.1mmol/L increments.
Adjustable settings in CL ICR, active insulin time, glucose target Basal rates, ICR, ISF, target range ICR, glucose target
Non-adjustable in CL Basal rates, ISF (automatically calculated and adapted) Active insulin time (set at 5 hours) Basal rates, active insulin time, ISF (all automatically calculated and adapted)
Exercise mode 670G: No780G: Yes Yes Yes
Boost mode No No Yes
Revert Sick day rules —————————————–Recommended to revert to open loop for illness and/or ketones—————————————-
Automatically reverts to open loop if… Prolonged hyperglycemia (670G only), max/min insulin delivery, loss of CGM data, sensor integrity concerns, lack of calibrations. Loss of CGM data Loss of CGM data or loss of pump connectivity
Educate Meal bolus ———————————————-Pre-meal bolusing recommended for optimal outcomes———————————————
Late bolusing can lead to insulin stacking and hypoglycaemia as CL insulin delivery increases in response to rising glucose
Hypo treatment ——————Consider treating hypoglycaemia with fewer carbohydrates depending on recent insulin delivery——————-
System optimisation • System requires finger pricks for HCL functioning
• Use of temp target will turn off automated corrections (780G only)
• Extended bolus / combo bolus function not available in CL
• Set sleep activity schedule overnight for tighter target
• Adjust doses for individuals with shorter active insulin time
• Extended bolus possible in CL, max 2 hours

• Use exercise mode “Ease-Off” following hypoglycaemia
• Use “Boost” mode during periods of high glucose
• Extended bolus / combo bolus function not available in CL
Online training No Yes Yes
Sensor Type of sensor Guardian 3 Dexcom G6 Dexcom G6
Calibrations 670G: 4–6 per day780G: 1–2 per day Rarely required (factory calibrated) Rarely required (factory calibrated)
Sensor life 7 days 10 days 10 days
Share Remote monitoring No (App in development for 780G) Yes – Dexcom follow Yes – Text Alert (Dexcom Follow planned for 2021)
Upload/Data sharing Manual downloading Automated cloud storage for Dexcom data; manual downloading for pump Automated cloud storage to Diasend (Clarity in 2021)
Remote bolusing No No Yes

MPC, model predictive control; CGM, continuous glucose monitor; TDD, total daily insulin dose; ICR, insulin-to-carbohydrate ratio; ISF, insulin-sensitivity factor; CL, closed-loop.