Failure rate of 600 pulse oximeter measurements with a transmittance pulse oximeter probe placed on the tongue and a reflectance pulse oximeter probe placed on the tail or tibia in 100 anesthetized dogs.

Total Transmittance probe tongue Reflectance probe tail Reflectance probe tibia P-value
n Failure rate n Failure rate n Failure rate n Failure rate Tongue vs. tail Tongue vs. tibia Tail vs. tibia
BUT 600 14.3% 300 3.7% 150 12.0% 150 38.0% 0.0007 <0.0001 <0.0001
DEX 600 27.5% 300 12.5% 150 28.0% 150 57.3% <0.0001 <0.0001 <0.0001

n, number of measurements; BUT, premedication butorphanol 0.2 mg/kg IV; DEX, premedication dexmedetomidine 5 μg/kg + butorphanol 0.2 mg/kg IV.

Measurement failure was defined as: pulse rate or SpO2 not displayed, signal intensity ≤2 of 10 signal intensity bars, or pulse rate differed from ECG heart rate by >10 beats per minute. Bold values mean significant differences.