Table 3.

Bland-Altman analysis of the pulse oximetry values displayed by a transmittance probe placed on the tongue and a reflectance probe placed on the base of the tail in anesthetized dogs.

Tongue vs. tail All BUT DEX
BIAS 95% LoA BIAS 95% LoA BIAS 95% LoA
SpO2 (%) −2.2 ± 3.8 −9.7–5.2 −3.1 ± 2.6 −8.3–2.0 1.0 ± 4.7 −8.2–10.2
PR (/min) −0.1 ± 3.2 −6.1–6.1 −0.3 ± 2.4 −5.0–4.5 0.0 ± 4.0 −7.8–7.8
Tongue vs. tibia
SpO2 (%) −0.6 ± 7.2 −14.8–13.6 −2.0 ± 6.8 −15.5–11.4 1.6 ± 7.3 −12.8–16.0
PR (/min) −0.2 ± 2.7 −5.2–5.5 −0.1 ± 2.9 −5.7–5.5 0.5 ± 2.4 −4.3–5.3
Tail vs. tibia
SpO2 (%) −2.2 ± 6.9 −11.4–15.7 −0.8 ± 4.5 −8.1–9.8 4.4 ± 9.4 −14.0–22.8
PR (/min) −3.7 ± 10.8 −24.8–17.5 −3.3 ± 10.3 −23.5–16.8 4.2 ± 11.7 −27.1–18.7

All, all dogs; BUT, dogs premedicated with butorphanol 0.2 mg/kg IV; DEX, dogs premedicated with dexmedetomidine 5 μg/kg IV and butorphanol 0.2 mg/kg IV; 95% LoA, 95% limits of agreement; SpO2, oxygen saturation measured by pulse oximetry; PR, pulse rate measured by pulse oximetry; SQ, signal quality measured by pulse oximetry; PI, perfusion index measured by pulse oximetry.