Table 3.

Intra-batch accuracy and precision, and inter-batch accuracy and precision using liquid chromatography–mass spectrometry.

Analyte Nominal concentration (ng/mL) Day 1
Day 2
Day 3
Inter-day
Accuracy* (%) Precision (%) Accuracy (%) Precision (%) Accuracy (%) Precision (%) Accuracy (%) Precision§ (%)
α-TP 2.50 114 6.58 119 2.50 109 5.00 114 5.83
250 99.1 5.23 115 3.63 101 6.14 105 7.96
800 103 4.95 108 2.23 102 4.44 104 4.55
γ-TP 2.50 102 17.5 81.4 9.63 103 5.67 95.6 15.8
250 113 7.52 76.9 4.53 82.6 6.99 93.7 19.6
800 101 8.57 79.3 21.5 84.0 7.80 88.2 16.5
α-TT 2.50 84.3 5.10 80.4 9.97 92.6 7.95 85.8 9.53
250 94.6 7.98 109 4.39 109 7.12 104 9.11
800 91.9 14.9 107 2.03 109 1.58 103 10.6
γ-TT 2.50 115 8.06 93.1 18.7 106 5.21 104 13.7
250 75.7 6.63 76.5 22.1 73.9 11.5 75.3 14.0
800 78.0 15.3 87.7 11.4 77.2 3.29 80.9 12.0
α-CMBHC 2.50 99.5 7.11 116 2.86 98.8 24.4 105 8.88
250 98.3 5.60 116 0.96 98.9 11.8 105 10.6
800 101 12.5 109 3.14 107 2.87 106 7.57
α-CEHC 2.50 114 5.65 118 2.88 110 4.12 114 10.2
250 108 5.46 118 2.33 116 4.18 114 5.48
800 101 11.5 110 1.96 112 2.81 108 7.46
γ-CEHC 2.50 114 2.87 124 3.27 115 5.48 118 5.29
250 117 1.88 121 3.55 113 6.80 117 5.07
800 109 6.02 108 2.47 118 6.09 111 6.20
*

Intra-day accuracy was measured as the amount of target analytes added to a blank matrix, carried throughout the extraction procedure, and then analyzed. The % accuracy was based on the mean of 5 samples at low, medium, and high concentration levels.

Inter-day precision was measured as the amount of target analytes added to a blank matrix, carried throughout the extraction procedure, and then analyzed. The inter-day % precision is determined by the average % recoveries obtained in days 1–3.

Intra-day accuracy was measured by the amounts obtained from the coefficients of variation (% CV). The % accuracy is based on the averages of the % CVs of 5 samples at low, medium, and high concentration levels.

§ Inter-day precision was measured by the amounts obtained from the coefficients of variation (% CV). The inter-day precision % is determined by the average % precisions obtained in days 1–3.