Skip to main content
. 2020 Dec 30;44(5):zsaa291. doi: 10.1093/sleep/zsaa291

Table 5.

Sleep summary: light sleep

Device n PSG
Mean ± SD
Device
Mean ± SD
Bias Lower limit Upper limit t (p) Effect size R 2 (p)
Fitbit Alta HR 49 236.6 ± 28.5 256.7 ± 30.1 20.0 –54.1 94.2 3.8 (<0.001) 0.68 0.00 (0.714)
Garmin Fenix 5S 29 238.3 ± 36.2 267.3 ± 35.8 29.0 –74.4 132.4 3.0 (0.005) 0.80 0.00 (0.957)
Garmin Vivosmart 3 43 238.3 ± 31.9 273.0 ± 36.1 34.7 –60.5 129.8 4.8 (<0.001) 1.01 0.02 (0.431)
Earlysense Live 51 237.3 ± 31.0 215.0 ± 51.5 –22.3 –133.6 89.0 –2.9 (0.006) –0.52 0.22 (<0.001)
ResMed S+ 51 235.9 ± 31.1 253.0 ± 34.3 17.1 –58.5 92.6 3.2 (0.002) 0.52 0.01 (0.468)
SleepScore Max 42 236.5 ± 32.2 259.1 ± 38.7 22.7 –51.7 97.0 4.0 (<0.001) 0.63 0.04 (0.196)

Summary results for total minutes of light sleep for the devices versus polysomnography (PSG). For PSG, light sleep was calculated as the combination of N1 and N2 sleep stages. Results are shown for all devices that output sleep stage classifications. See Table 1 caption for additional table details.