. 2021 Apr 29;16(2):271–286. doi: 10.18502/jovr.v16i2.9091

Table 1.

The prevalence of optical coherence tomography angiography artifacts


Authors Design Studied group Artifacts reported Device type and Software version
Chen et al, 2016[19] Prospective observational study Normal Retinal diseases Glaucoma Motion (100%) Fringe washout (100%) Projection (100%) Masking and unmasking (100%) Stromal decorrelation signal (100%) Retinal vessels duplication (5%) RTVue XR Avanti system (Optovue Inc., CA, US)
Ghasemi Falavarjani et al, 2016[13] Retrospective observational study Age-related macular degeneration Diabetic retinopathy Retinal vascular occlusions Band (89.4%) Segmentation (61.4%) Motion (49.1%) Unmasking (15.8%) Blink (8.8%) Doubling of the retinal vessels (1.7%) Out of window (1.7%) Masking (1.7%) Projection (0.0%) Stretch artifacts (0.0%) Crisscross (0.0%) Topcon OCT instrument (DRI OCT Triton plus, Topcon, Tokyo, Japan).
Al-Sheikh et al, 2017[23] Prospective comparative study Healthy subjects Band (17.64–70.58%) Segmentation (5.8–11.6%) Motion (5.8%) Projection (0–47.05%) DRI OCT Triton, TOPCON Inc., Tokyo, Japan
Say et al, 2017[20] Observational study Treated unilateral posterior uveal melanoma and fellow eye Loss of focus (55%) Broad blink lines (37%) Motion (26%) Specular dot (25%) Edge duplication (8%) The Optovue RTVue XR AVANTI, version 2014.2.0.13
Holmen et al, 2019[15] Cross-sectional study Diabetic retinopathy Eye movement (93.1%) Defocus (74.9%) Shadow (62.3%) Tilt (50.5%) Z offset (43.8%) Refraction shift (31.8%) Segmentation (24.6%) Decentration (21.4%) Projection (6.7%) Blink (1%>) Stretch artifact (1%>) Edge duplication (1%>) Loss of signal (1%>) CIRRUS HD-OCT 5000, with the AngioPlex module, version 10.0.0.13424 or the Optovue Avanti RTVue XR, version 2018.0.0.5).
Enders et al, 2019[21] Prospective observational study Healthy subjects Neovascular age-related macular degeneration Diabetic retinopathy Retinal vascular occlusions Projection (100%) Segmentation (55%) Motion (49%) Masking (45%) Band (43%) Blink (28%) Window artifact (8%) Vessel doubling (0.0%) Stretch artifacts (0.0%) Spectral domain OCT Cirrus 5000 equipped with the AngioPlex module)
Iftikhar et al, 2019[16] Prospective cross-sectional study Multiple Sclerosis Healthy subjects Motion (96.3%) Blink (51.9%) Loss of focus (25.1%) SpectralisⓇ OCT-A prototype, OCT2 (Heidelberg, Germany)
Eastline et al, 2019[68] Prospective observational study Healthy eyes Displacement artefact (96.34%) Shadowing (92.27%) White line artefacts (63.41%) Vessel doubling (35.37%). PLEX Elite 9000 scanner (Carl Zeiss Meditec, Inc., Dublin, CA). central 3 * 3 and 12 * 12 mm scans of the four quadrants (wide field)
Weijing et al, 2020[17] Retrospective observational study Glaucomatous eyes Healthy eyes Projection (100%) Motion (75.22%) Blink (2.62%) Stretching (0.87%) Blurred images (24.78%) Decentration (21.28%) Vignetting (2.04%) Unmasking (0.87%) Segmentation (1.17%) Out of window (0.29%) Vessel doubling (0.58%) N/A
Bontzos et al, 2020[69] Prospective observational study Eyes with idiopathic Epiretinal membranes Healthy eyes Segmentation (0% in healthy eyes, 64.1% in ERM patients) Motion (7.5 % in healthy eyes, 53.8 % in ERM patients) AngioVue, RTVue XR Avanti SD-OCT, Optovue Software version: 2016.2.0.35