Table 3.

Gamma pass rates for simulated treatment plans comparing the motion compensated dose distributions to planned. All dose differences are relative to the global maximum dose. A dose threshold of 10% of the prescription dose (50 Gy) is applied to all data. Gamma analyses are 3D. The extreme locations correspond to shifts in IEC‐Y at isocenter of ± 2.0 cm and ± 1.25 cm for the 1.0‐cm and 2.5‐cm jaws, respectively.

Jaw Position Fixed‐Width Synchrony
3%, 2 mm a 1%, 1 mm 3%, 2mm a 1%, 1 mm
2.5 cm Negative extreme 100 94.67 100 99.20
Positive extreme 100 94.44 100 98.70
1.0 cm Negative extreme 62.16 48.11 100 98.89
Positive extreme 62.01 47.02 100 99.05
a

Universal tolerance limit recommended by TG‐218 for IMRT QA is > 95% of points passing 3%, 2 mm, with a 10% dose threshold. 10