Table 2.

Much larger inter‐fractional CTV displacement measured by CBCT than the intra‐fractional one in PMRM‐IMRT for left‐sided breast cancer during peaceful breathing.

Paramaters Direction Number Mean ± SD(mm) ≤2 mm ≤3 mm ≤4 mm
Intra‐fraction X 900 0.3 ± 0.3 99.7% 100.0% 100.0%
Y 900 0.5 ± 1.0 88.1% 96.5% 100.0%
Z 900 0.5 ± 0.5 96.5% 99.5% 100.0%
Sum 2700 0.4 ± 0.7 94.8% 98.7% 100.0%
Inter‐fraction X 30 1.7 ± 1.2 40.0% 93.3% 96.6%
Y 30 2.8 ± 2.3 46.6% 60.0% 63.3%
Z 30 2.6 ± 2.2 50.0% 66.6% 70.0%
Sum 90 2.4 ± 2.1 45.5% 73.3% 76.6%

X, Y, and Z values indicated the inter‐ or intra‐fractional CTV displacement in x (laterally), in y (longitudinally) and in z coordinate axis (vertically), respectively. Data presented as mean ± standard deviation (Mean ± SD) (mm). The values in the last three columns were the CTV coverage rates under the corresponding status when using CTV expanding values with 2, 3, and 4 mm, respectively. Abbreviations: PMRM‐IMRT = post modified radical mastectomy intensity‐modulated radiotherapy. CBCT = cone beam computed tomography.