Table 1.
C57BL/6 brain (mouse) | Spinal cord (mouse) | V1 cortex (human) | CC splenium/body (human) | Rag2−/−(sh−/−) brains (mouse) | |
---|---|---|---|---|---|
Beamline, synchrotron | cSAXS, SLS | LiX, NSLS-II | 4-2, SSRL | cSAXS, SLS | LiX, NSLS-II |
Tube diameter (mm) | 10 | 3 | 20a | 10 | 10 |
Beam energy (keV) | 16.3 | 15.7 | 17 | 16.3 | 16.9 |
Beam size (μm2) | 150 × 150 | 100 × 100 | 250 × 250 | 150 × 150 | 120 × 120 |
Motor step size (μm) | 150 | 100 | 250 | 150 | 120 |
Field of view (pixels) | 69 × 105 | 38 × 61 | 85 × 90 | 70 × 82 | 91 × 41 |
Field of view, size (mm2) | 10.35 × 15.75 | 3.8 × 6.1 | 21 × 22.5a | 10.5 × 12.3 | 10.92 × 4.92 |
No. of projections | 267 | 60 | 151 | 254 | 181 |
Rotation axis tilt angles (°) | 0, 10, 15, 20, 30, 40, 45, 50 | 0 | 0 | 0, 10, 20, 30, 40, 50 | 0 |
Nr of projections/tilt | 67, 34, 32, 32, 30, 26, 24, 22 | 60 | 151 | 66, 56, 50, 44, 38 | 181 |
Exposure time/frame (ms) | 125 | 300 | 120 | 80 | 100 |
Total exposure time (h) | 67.1 | 11.6 | 38.5a | 25.3 | 18.8 |
Total scan time (h) | 87.2 | 17.4 | 61a | 36.8 | 24.8 |
Doseb (Gy) | 2.33 × 104 | 9.96 × 104 | 1.1 × 104 | 1.38 × 104 | 2.48 × 104 |
Reconstruction iterations | 10,000 | 2,000 | 5,000 | 5,000 | 5,000 |
a These values include scanning of two more specimens, not incorporated in the present study, that were scanned together with the V1 sample (in the same tube, next to one another, see Supplementary Fig. 1).
bDose in was calculated as