Table 2.

Studies with ultra-low-dose coronary CT angiography with mean effective radiation dose < 2.2 mSv.

Study Number of Patients Heart Rate/bpm BMI/ kg m-2 CT Scanner Dose Reduction Mean Effective Dose/ mSv
Yin et al. (2013) [81] 21 50 ± 6 23.9 ± 3.2 128 detector row, second-generation dual Source CT (Definition Flash, Siemens Healthcare, Forchheim, Germany). High-pitch spiral acquisition and raw data-based iterative reconstruction. Tube voltage was set to 80 kV and tube current was 50 mAs. 0.06 ± 0.01
Zhang et al. (2014) [82] 58 60.4 ± 5.7 22.5 ± 1.9 Dual-source CT system (Somatom Flash; Siemens Medical Solutions, Forchheim, Germany). Prospective ECG-triggered high-pitch spiral acquisition (3.4) at 70 kVp with 30 mL of contrast agent. Automated tube current modulation (CAREDose 4D, Siemens). Image reconstruction with iterative reconstruction SAFIRE (SAFIRE, Siemens, strength-level 3). 0.17 ± 0.02
Zhang et al. (2016) [83] 43 69.4 ± 13.6 23.3 ± 2.8 64-slice, second-generation dual-source CT system (Somatom Flash; Siemens Medical Solutions, Forchheim, Germany). Prospective ECG-triggered high-pitch spiral acquisition (3.4) at 70 kVp. Automated tube current modulation (CAREDose 4D, Siemens) was enabled. 0.20 ± 0.00
Stehli et al. (2014) [84] 36 57.6 ± 6.2 27.6 ± 4.7 64-slice CT scanner (Discovery HD 750, GE Healthcare). Prospective ECG triggering. Body mass index (BMI)–adapted tube voltage and tube current. Images reconstructed using MBIR algorithms. 0.29 ± 0.12
Hell et al. (2014) [85] 26 54 ± 5 27.7 ± 3.8 192-slice, third-generation dual-source CT system (Somatom Force; Siemens Healthcare, Forchheim, Germany). A prospectively ECG-triggered high-pitch (3.2) spiral (flash) acquisition was performed.
Tube voltage was set at 70 kVp and tube
current at 450 mAs. Images were reconstructed using iterative algorithm ADMIRE (ADMIRE; Siemens Healthcare, Forchheim, strength level 2)
0.30 ± 0.03
Gordic et al. (2016) [86] 25* 61 ± 5 25.3 ± 3.4 192-slice dual-source CT (SOMATOM Force, Siemens Healthcare). Prospective ECG-triggered, high-pitch spiral acquisition. Automated attenuation-based tube voltage selection (CAREkV, Siemens) and tube current modulation (CAREDose, Siemens) was applied. In Images reconstructed with ADMIRE (ADMIRE; Siemens Healthcare, Forchheim, strength level 4). 0.3 ± 0.1
Wang et al. (2012) [69] 40* 55.4 ± 4.8 20.6 ± 1.4 Dual source CT scanner (Definition Flash, Siemens AG, Forchheim, Germany). Prospective ECG-triggering high-pitch spiral (3.4) at 80 kV. 0.36 ± 0.03
Yin et al. (2013) [81] 40 54 ± 4 25.5 ± 3.1 64-slice, second-generation dual-source CT system (Somatom Definition Flash, Siemens Healthcare, Forchheim, Germany). Prospective ECG-triggered high-pitch spiral acquisition (3.4). Attenuation-based tube current modulation (CareDose 4D, Siemens). 0.58 ± 0.17
Neefjes
et al. (2011) [87]
80* 58 ± 7 28 ± 4 64-slice, dual-source CT scanner (Somatom Definition Flash; Siemens Healthcare, Forchheim, Germany). Prospective ECG-triggered high-pitch spiral (3.4) 0.81 ± 0.3
Study Number of Patients Heart Rate/bpm BMI/ kg m-2 CT Scanner Dose Reduction Mean Effective Dose/ mSv
Achenbach et al. (2010) [70] 50 68 ± 9 25.4 ± 4.8 128 slice, dual-source CT system (‘Definition Flash’, Siemens Healthcare, Forchheim, Germany) Prospective ECG-triggered high-pitch spiral acquisition ((pitch was 3.2 in first 28 patients and 3.4 in last 22 patients. Tube voltage was 100 kV and tube current was 320 mA s/rot. 0.87 ± 0.07
Alkadhi
et al. (2010) [30]
50* 56 ± 10 25.9 ± 2.8 128-slice, second-generation dual-source CT scanner (Somatom Definition Flash; Siemens Healthcare, Forchheim, Germany). Prospectively ECG-gated high-pitch mode (3.4). 0.9 ± 0.1
Chen et al. (2013) [53] 107* 57.1 ± 11.2 27.3 (24.6–32.3) 320 row, second-generation CT scanner (Aquilion One Vision Edition; Toshiba Medical Systems, Otawara, Japan). Prospectively ECG-gated. Tube potential and tube current were determined with use of automatic exposure control (SURE Exposure3D, Toshiba Medical Systems). Image reconstruction with iterative reconstruction AIDR-3D ((Toshiba Medical Systems). 0.93
Stolzmann et al. (2011) [42] 100 66 ± 20 27.7 ± 4.2 Second-generation dual-source 128-MDCT CT scanner (Somatom Definition Flash, Siemens Healthcare). Prospective ECG-triggered high-pitch spiral acquisition (3.4). Tube voltage 100 kVp and tube current-time product 320 mAs per rotation 1.0 ± 0.2
Lell et al. (2009) [48] 25 69 ± 9 26.8 ± 5.6 64 detector row, dual source CT system (Definition Flash, Siemens Healthcare, Forchheim, Germany). Prospective ECG-triggered high-pitch spiral acquisition (3.2). Tube settings were 100 kV/320 mAs and 120 kV/400 mAs for patients below and above 100-kg weight, respectively. 1.0 ± 0.3
Sun et al. (2012) [57] 134* 79 ± 9 23.8 ± 2.7 Second-generation DSCT system (Somatom Definition Flash, Siemens Healthcare, Forchheim, Germany). Prospective ECG-gated high-pitch spiral acquisition (3.4). Tube voltage 100 kVp and tube current-time product 320 mAs per rotation 1.04 ± 0.16
Sommer
et al. (2010) [49]
33 57.3 ± 7.0 24.7 ± 2.0 Second generation dual source CT scanner (Somatom Definition Flash, Siemens Healthcare). Prospective ECG-triggered high-pitch spiral acquisition (3.4). 1.11 ± 0.14
Wang et al. (2014) [80] 172* 59.8 ± 7.6 24.2 ± 2.5 Dual-source CT scanner (Somatom Definition Flash; Siemens, Munich, Germany). Prospective ECG-triggered high-pitch helical mode (flash mode) d if the patient’s HR < 65 bpm and prospectively ECG-triggered sequential mode if the patient’s HR between 65 and 90 bpm. Automatic tube potential selection (CARE kV; Siemens) and automatic tube current selection algorithm (CARE Dose4D; Siemens). 1.25 ± 1.24
Wichmann et al. (2015) [47] 25* 53 ± 2 128-slice second-generation dual-source CT (Somatom Definition Flash, Siemens Healthcare). Prosepctive ECG-gated high-pitch spiral. Automatic tube potential selection (CARE kV; Siemens) and automatic tube current selection algorithm (CARE Dose4D; Siemens). 1.27 ± 0.62
Huang
et al. (2015) [34]
70* 58.6 ± 4.5 24.8 ± 2.8 64-slice DSCT scanner (Somatom Definition, Siemens Healthcare, Forchheim, Germany). Single-phase, prospective ECG-triggered acquisition. 1.27 ± 0.57
Study Number of Patients Heart Rate/bpm BMI/ kg m-2 CT Scanner Dose Reduction Mean Effective Dose/ mSv
Koplay
et al. (2016) [88]
186 66.52 ± 11 27.97 (19–40) 128 slice dual-source CT (Somatom Definition Flash, Siemens, Germany). Prospective ECG-triggered high-pitch spiral acquisition. Automatic tube potential selection (CARE kV; Siemens) and automatic tube current selection algorithm (CARE Dose4D; Siemens). 1.3 ± 0.4
Leipsic
et al. (2011) [64]
24* 54 ± 5 27 ± 4 Discovery HD 750 (GE Healthcare, Waukasha, WI, USA). Prospective ECG-triggering. Reduced tube voltage was defined as 80 or 100 kVp for individuals with BMI < 25 kg/m2 or 25–35 kg/m2, respectively; whereas standard tube voltage was defined as 100 or 120 kVp for individuals with BMI < 25 kg/m2 or 25–35 kg/m2, respectively. 1.3 ± 0.5
Matsubara et al. (2016) [45] 17* 59.1 ± 6.0 21.0 ± 2.0 128-slice dual-source CT Somatom Definition Flash scanner (Siemens Healthcare) Prospective ECG-triggered high-pitch spiral acquisition (3.4). 1.5 ± 0.2
Pflederer
et al. (2010) [89]
56* Dual-source CT (Definition, Siemens Healthcare, Forchheim, Germany). Prospective ECG-triggering. Tube settings were 100 kV/330 mAs. 1.5 ± 0.4
Mangold
et al. (2016) [66]
43* 68.1 ± 18.4 23.5 ± 3.3 3rd generation dual-source CT (Somatom Force, Siemens Healthcare, Forchheim, Germany). Prospective ECG-triggered spiral acquisition. Automated tube current selection and advanced iterative reconstruction. Tube voltage at 70 kV group. 1.5 ± 1.2
Kim et al. (2011) [28] 23* 59 ± 7 24.3 ± 3.5 128-slice MDCT (Definition AS Plus 128; Siemens, Forchheim, Germany). Step-and-shoot prospective ECG-gated group. 1.75 ± 0.83
Leipsic
et al. (2013) [90]
109* 59 ± 4 27.23 ± 4.27 Discovery HD 750 (GE Healthcare, Waukasha, WI, USA) and a Toshiba Aquilion One (Toshiba Medical Systems, Tokyo, Japan). Prospective ECG-triggering with a narrow window acquisition window. 1.78
Deseive
et al. (2015) [46]
150* 55.5 ± 5.0 26.1 ± 3.5 Second-generation dual-source CT scanner (Somatom, Definition Flash scanner, Siemens Medical Solutions) PROTECTION IV study. Prospective ECG-gated high-pitch helical acquisition. 100-kV tube potentials up to a body mass index of 30 kg/m2. 2.0 ± 2.4
Yoo et al. (2013) [78] 51 55 (39–65) 25.3 (18.8–32.8) 640-multi-slice CT scanner (Aquilion ONE; Toshiba Medical Systems, Tochiki-ken, Japan). Prospective ECG-triggering. Automatic exposure control system (SUREExposure; Toshiba Medical Systems, Tochiki-ken, Japan). Image reconstruction with AIDR-3D (standard). 2 ± 1
Duarte
et al. (2010) [91]
40* 60 ± 5 Weight/kg (70 ± 10) 128-MDCT (Somatom Definition AS128, Siemens Medical Solutions, Germany). Prospective ECG-gating with full tube current at 70%. Automated tube current to patient-specific parameters such as size and attenuation of body region (CAREDose system). 2.1 ± 0.9
Husmann
et al. (2009) [92]
100 57 ± 6 27 ± 4 64-slice, LightSpeed VCT XT scanner (GE Healthcare). Prospective ECG-triggering. 2.2 ± 0.7