Skip to main content
. 2021 May 24;23:62. doi: 10.1186/s12968-021-00751-2

Table 2.

Average contrast ratio (CR) and contrast-to-noise (CNR) ratio measured between blood-myocardium, scar-myocardium and scar-blood for both 2D and 3D LGE PSIR images

2D 3D
CR CNR CR CNR
Blood-myocardium 1.29 ± 0.08 4.85 ± 2.01 1.28 ± 0.16 6.03 ± 2.78
Scar-myocardium 1.43 ± 0.15 7.02 ± 2.4 1.36 ± 0.12 8.45 ± 3.16
Scar-blood 1.15 ± 0.09 2.79 ± 1.29 1.14 ± 0.07 4.15 ± 2.48