. 2021 Apr 22;11(5):746. doi: 10.3390/diagnostics11050746

Table 2.

Overall sensitivity and specificity of the Genedrive® HCV ID Kit compared with the sensitivity and specificity of the Abbott RealTime HCV Viral Load Assay at different detection thresholds.

Abbott: 12–30 IU/mL
Threshold
Total Diagnostic Accuracy (95% CI)
Target
Detected
Target
Undetected
Genedrive® HCV ID assay Positive 205 1 206 Sensitivity: 96.2% (92.7–98.4)
Specificity: 99.5% (97.4–100)
Negative 8 212 220
Total 213 213 426
Abbott: 1000 IU/mL threshold Total
Target detected Target undetected
Genedrive® HCV ID assay Positive 205 1 206 Sensitivity: 100% (98.2–100)
Specificity: 99.5% (97.5–100)
Negative 0 220 220
Total 205 221 426
Abbott: 2362 IU/mL threshold Total
Target detected Target undetected
Genedrive® HCV ID assay Positive 203 3 206 Sensitivity: 100% (98.2–100)
Specificity: 98.7 (96.1–99.7)
Negative 0 220 220
Total 203 223 426

Abbreviations: IU: International Units; CI: Confidence interval.