Table 1.

Studies before 1992 comparing SpO2 monitoring to invasively measured blood oxygenation parameters.

Ref 1st Author, Year Number of Patients/ HbF Blood Samples Blood Sample Type HbF Measurement Method Gestation Distribution (Weeks) Time of Sample Collection and Non-Invasive Monitoring Blood Oxygenation Parameters Blood Gas Analyzer /Hemoximeter Pulse Oximeter (Company Name) Additional Bedside Oxygenation Monitoring Device (Company Name) Relevant Results
[32]. Durand,
1986
75/140 Arterial Alkali denaturation method 24–42 1–14 days + 30–153 days after birth paO2, SaO2 Radiometer BMS3 Mark II / Co-oximeter IL 282 Nellcor N-100 (Hayward, CA, USA) tc-pO2 Oxygen electrode (Novametrix, Wallingford, CT, USA) HbF values of 4.3% to 95% did not influence the accuracy of pulse oximeter readings.
[7]. Ramanathan, 1987 68/132 Arterial Alkali denaturation method 25–31 1–6 days + 20–80 days after birth paO2, SaO2 Radiometer BMS3 Mark II / Co-oximeter IL 282 Nellcor N-100 (Hayward, CA, USA) tc-pO2 Oxygen electrode (Novametrix, Wallingford, CT, USA) HbF values of 4.3% to 92.2% did not influence the accuracy of pulse oximeter readings.
[33]. Wimberley, 1987 18/18 Arterial Alkali denaturation method 25–34 Within 5 days after birth paO2, SaO2 ABL300/ Hemoximeter OSM3 Ohmeda Biox 3700 tc-pO2 Radiometer TCM3 FHbF ranged from 44–97%. The variations in the levels of HbF, pH, pCO2 and 2,3-DPG resulted in a variable paO2-SaO2 relation.
[4]. Jennis,
1987
26/49 Arterial Electrophoresis 24–40 1–49 days after birth SaO2 Co-oximeter IL-282 Nellcor N-100 (Hayward, CA, USA) NA FHbF > 50% generated a 2.8% to 3.6% error (underestimation) in SpO2 reading.
[5]. Praud,
1989
71/52 Arterial Electrophoresis and alkali denaturation method 25–40 1–14 days after birth + 4.5–38 weeks after birth SaO2 Hemoximeter OSM2 Nellcor N-100 (Hayward, CA, USA) NA For FHbF < 50% and SaO2 ≤ 95%, SpO2 was overestimated.

FHbF = fraction of fetal hemoglobin, HbF = fetal hemoglobin, NA = not applicable, paO2 = partial arterial oxygen pressure, SaO2 = arterial blood oxygen saturation, SpO2 = peripheral arterial oxygen saturation measured by pulse-oximetry, tc-pO2 = transcutaneous oxygen tension, 2,3-DPG = 2,3- diphosphogylcerate.