Table 4.

(A) Two by Two table for diagnostic performance of mAb 18B7 ICT strip against serum samples compared with the rapid test of CrAg® IMMY; (B) Summary of diagnostic performance of mAb 18B7 ICT strip with serum samples.

(A)
mAb 18B7 ICT Strip Rapid Test of CrAg® IMMY Total
Cryptococcosis
(Positive)
Non-Cryptococcosis
(Negative)
Positive 50
(TP) *
1
(FP) *
51
Negative 2
(FN) *
44
(TN) *
46
Total 52 45 97
(B)
Diagnostic Performance Criteria for Serum Percentages (95% C.I.)
Sensitivity 96.15% (86.28–99.68)
Specificity 97.78% (87.37–99.94)
Positive predictive value 98.04% (88.73–99.95)
Negative predictive value 96.65% (85.16–99.47)
Accuracy 96.91% (91.23–99.36)
Cohen’s kappa coefficient (κ) 0.938 (0.869–1.000)

* TP, True-Positive; TN, True-Negative; FP, False-Positive; FN, False-Negative.