Skip to main content
. 2021 Apr 29;13(5):796. doi: 10.3390/v13050796

Table 1.

Results of the test using the GenBody COVID-19 Ag test (COVAG025).

Evaluation Results of Test Equipment (GenBody COVID-19 Ag Test (COVAG025) Confirmed Results through RT-PCR Total
Positive Negative
Positive 27 2 29
Negative 3 98 101
Total 30 100 130

NB. Clinical sensitivity: 90.00% (27/30), (95% CI: 73.47–97.89%); Clinical specificity: 98.00% (98/100), (95% CI: 92.96–99.76%); Positive predictive value: 93.10% (27/29), (95% CI: 77.23–99.15%); Negative predictive value: 97.03% (98/101), (95% CI: 91.56–99.38%); Accuracy: 96.15% (125/130), (95% CI: 91.25–98.74%).