. 2021 May 26;36(3):879–888. doi: 10.1007/s10877-021-00721-0

Table 3.

Cardiac index measurements by FloTrac compared to bolus thermodilution technique with a pulmonary artery catheter

FloTrac All
n = 579
Phase 1
n = 24
Phase 2
n = 27
Phase 3
n = 85
Phase 4
n = 104

Bias

(L min−1 m−2)

(95% CI)

0.01

(− 0.05 to 0.06)

− 0.17

(− 0.44 to 0.10)

− 0.01

(− 0.27 to 0.26)

− 0.15

(− 0.30 to 0.00)

− 0.25

(− 0.39 to − 0.11)

LOA lower

(L min−1 m−2)

(95% CI)

− 1.27

(− 1.56 to − 0.98)

− 1.28

(− 1.8 to − 0.77)

− 1.32

(− 1.85 to − 0.78)

− 1.52

(− 1.99 to − 1.06)

− 1.78

(− 2.37 to − 1.18)

LOA upper

(L min−1 m−2)

(95% CI)

1.29

(1.0 to 1.58)

0.93

(0.41 to 1.45)

1.26

(0.72 to 1.79)

1.08

(0.61 to 1.54)

1.27

(0.67 to 1.87)

Percentage error

(95%CI)

56.8%

(44.4 to 69.2)

47.2%

(32.8 to 61.6)

53.7%

(32.7 to 74.7)

65.7%

(48.1 to 83.3)

61.9%

(30.4 to 93.4)

Regression coefficient

(L min−1 m−2)

(95% CI)

− 0.19

(− 0.27 to − 0.10)

− 0.12

(− 0.67 to 0.43)

− 0.26

(− 0.98 to 0.47)

− 0.06

(− 0.37 to 0.26)

− 0.31

(− 0.52 to − 0.09)

Error grid
 Zone 1 39.3% 36.8% 45.5%
 Zone 2 12.2% 11.8% 8.0%
 Zone 3 35.4% 31.6% 38.6%
 Zone 4 13.0% 19.7% 8.0%

Phase 1 is before induction of anesthesia, phase 2 is during the construction of the proximal anastomoses to the ascending aorta, phase 3 is distal coronary anastomoses, phase 4 is postoperatively in the intensive care unit before extubation. Error grid analysis was not made for the phases 1 and 2 because of the small number of measurements. n number of CI measurements during the phase. LOA limits of agreement