Skip to main content
. 2021 May 13;21(10):3392. doi: 10.3390/s21103392

Table 1.

Performance comparison between our laser gas analyzer and a commercial instrument.

Our TDLAS Analyzer LI-7500 Analyzer
Method Laser absorption spectroscopy Non-dispersive infrared spectroscopy
Detection limit, H2O 3.25 ppmv at 10 Hz
8.17 ppmv at 100 Hz
4.70 ppmv at 10 Hz
6.70 ppmv at 20 Hz
Detection limit, CO2 0.13 ppmv at 10 Hz
0.40 ppmv at 100 Hz
0.11 ppmv at 10 Hz
0.16 ppmv at 20 Hz
Maximum data rate 100 Hz 20 Hz