Table 3.

Pairwise comparison between the e-CoVig and the standard device measurements.

Standard Device e-CoVig Device Mean Difference p-Value
Heart rate (BPM) 80.10±15.60 80.50±15.70 0.40±2.10 0.912
SpO2 (%) 97.50±2.50 97.40±2.30 0.10±0.70 0.829
Temporal temperature (°C) 36.37±0.33 36.71±0.24 0.34±0.34 <0.001
Tympanic temperature (°C) 36.62±0.18 35.57±0.23 1.04±0.26 <0.001