Table 10.
Inhibition Zone Diameter (mm) | ||||
---|---|---|---|---|
Incubation Duration (h) | Standard Conditions a | At the Alert | 18 h after the Alert, Incubation in the Device |
3 h after the Alert, at Room Temperature |
18 | 35–40 | 35–40 | 37–39 | 40–42 |
20 | 36–40 | 36–40 | 37–42 | 38–42 |
22 | 39–41 | 36–42 | 37–42 | 42 |
24 | 39–42 | 42 | 37–43 | 43–45 |
a Ciprofloxacin inhibition zone diameter for the standard culture were taken from Table 7; the diameters were determined by at least three independent assays.