Table 1. Quantitative Image Analysis.

40-keV Images Conventional Images P
CT attenuation of tumor (HU) 46.4 ± 8.4 34.6 ± 5.4 < 0.001
CT attenuation of WM (HU) 29.2 ± 4.9 26.2 ± 2.8 0.001
CT attenuation of GM (HU) 45.6 ± 5.4 34.4 ± 1.9 < 0.001
Image noise (HU) 5.5 ± 1.4 5.8 ± 1.2 0.03
Contrast of tumor-WM (HU) 17.3 ± 9.0 8.4 ± 5.5 < 0.001
Contrast of tumor-GM (HU) 0.8 ± 7.6 0.2 ± 5.4 0.47
CNR of tumor-WM 3.4 ± 1.9 1.5 ± 1.0 < 0.001
CNR of tumor-GM 0.2 ± 1.4 0.1 ± 0.9 0.31

Data are shown as mean ± standard deviation. CNR = contrast-to-noise ratio, GM = grey matter, HU = Hounsfield unit, WM = white matter