Skip to main content
. 2021 May 26;8:141. doi: 10.1038/s41597-021-00916-9

Table 6.

Statistics of the “check” standard quality-control metrics as presented in Fig. 6 for the three instrument types.

Instrument δ18O precision [‰] δ18O accuracy [‰] δD precision [‰] δD accuracy [‰]
L2120-i 0.065 ± 0.028 0.023 ± 0.059 0.524 ± 0.163 0.151 ± 0.232
L2130-i 0.031 ± 0.017 −0.001 ± 0.035 0.301 ± 0.098 0.034 ± 0.146