. 2021 May 27;21:487. doi: 10.1186/s12879-021-06134-3

Table 3.

Performance evaluation of commercial HBeAg RDTs using a plate ELISA as a reference test

Diagnostic performance characteristics (95% confidence interval)
SD BIOLINE HBeAg (Alere) HBeAg serum rapid test (Creative Diagnostics) HBeAg Rapid Test (Biopanda Reagents)
Overall (n = 194)
 κ statistica 1.0 (1.0–1.0) 0.96 (0.92–1.0) 0.89 (0.80–0.98)
 Sensitivity (%) 27.7 (15.6–42.6) 72.3 (57.4–84.4) 53.2 (38.1–67.9)
 Specificity (%) 100 (97.5–100) 99.3 (96.3–100) 95.9 (91.3–98.5
Community study (n = 100)
 κ statistic a 1.0 (1.0–1.0) 1.0 (1.0–1.0) 0.71 (0.41–1.0)
 Sensitivity (%) 36.4 (10.9–69.2) 63.6 (30.8–89.1) 36.4 (10.9–69.2)
 Specificity (%) 100 (95.9–100) 100 (95.9–100) 98.9 (93.9–100)
 Positive predictive value (%) 100 (39.8–100) 100 (59–100) 80.0 (28.4–99.5)
 Negative predictive value (%) 92.7 (85.6–97.0) 95.7 (89.4–98.8) 92.6 (85.4–97.0)
 Diagnostic accuracy (%) 93.0 (86.1–97.1) 96.0 (90.1–98.9) 92.0 (84.8–96.5)
Hospital study (n = 94)
 κ statistica 1.0 (1.0–1.0) 0.95 (0.88–1.0) 0.92 (0.83–1.0)
 Sensitivity (%) 25.0 (12.1–42.2) 75.0 (57.8–87.9) 58.3 (40.8–74.5)
 Specificity (%) 100 (93.8–100) 98.3 (90.8–100) 91.4 (81.0–97.1)
 Positive predictive value (%) 100 (66.4–100) 96.4 (81.7–99.9) 80.8 (60.6–93.4)
 Negative predictive value (%) 68.2 (57.2–77.9) 86.4 (75.7–93.6) 77.9 (66.2–87.1)
 Diagnostic accuracy (%) 71.3 (61.0–80.1) 89.4 (81.3–94.8) 78.7 (69.1–86.5)

Abbreviations: AROC Area under the receiver operating curve, RDT rapid diagnostic test, ELISA enzyme linked immunosorbent assay

aCohens’ κ statistic demonstrates inter-rater agreement