FIGURE 6.
Plasma 217+tau assay has wide dynamic range and good dilution linearity. A, A 12‐point calibrant curve was run on the plasma p217+tau assay on five separate days, revealing consistent signal with a wide dynamic range (5–30,000 fg/mL). B, A panel of Alzheimer's disease (AD) serum and plasma samples (n = 4) were measured in the plasma p217+tau assay at dilutions of 1:2, 1:3, 1:4, 1:5, and 1:6. Dilution corrected concentrations were compared to 1:3 dilution, revealing all dilutions tested were within 20% of each other