Skip to main content
. 2021 May 19;21(10):3535. doi: 10.3390/s21103535

Table 2.

Statistics of the response characteristics of the P1–P6 surfactant sensors to Hyamine, given together with ±95% confidence limits.

Parameters Surfactant Sensor with Plasticizer
P1 P2 P3 P4 P5 P6
Slope, mV/decade of activity 58.78 ± 0.6 53.61 ± 1.1 56.61 ± 0.5 50.6 ± 1.0 50.67 ± 1.1 -/-
Standard error 0.4 0.6 0.3 0.7 0.7 -/-
Correlation coefficient, R2 0.9935 0.9993 0.9992 0.9964 0.9965 -/-
Detection limit, M 2.1 × 10−7 4.2 × 10−7 0.1 x 10−6 0.2 × 10−6 0.1 x 10−6 -/-
Useful conc. Range, M 3.2 × 10−3–3.9 × 10−7 3.2 × 10−3–5.9 x 10−7 3.2 × 10−3–1.2 × 10−6 3.2 × 10−3–1.6 × 10−6 3.2 × 10−3–1.2 × 10−6 -/-