Statistics of the response characteristics of the P1–P6 surfactant sensors to DMIC, given together with ±95% confidence limits.

Parameters Surfactant Sensor with plasticizer
P1 P2 P3 P4 P5 P6
Slope, mV/decade of activity 58.4 ± 0.6 59.6 ± 0.7 54.6 ± 0.7 55.1 ± 0.7 55.9 ± 0.5 32.9 ± 0.4
Standard error 0.2 0.4 0.2 0.4 0.7 0.6
Correlation coefficient, R2 0.9994 0.9997 0.9998 0.9997 0.9999 0.9994
Detection limit, M 2.1 × 10−7 4.2 × 10−7 5.7 × 10−7 0.9 × 10−6 0.1 × 10−6 2.9 × 10−6
Useful conc. Range, M 3.2 × 10−3–3.9 × 10−7 3.2 × 10−3–5.9 × 10−7 3.2 × 10−3–7.9 × 10−7 3.2 × 10−3–2.6 × 10−6 3.2 × 10−3–1.2 × 10−6 3.2 × 10−3–4.9 × 10−6