Skip to main content
. 2021 May 17;12:664477. doi: 10.3389/fmicb.2021.664477

TABLE 1.

Results of the BioFire COVID-19 using serial dilutions of synthetic SARS-CoV-2 RNA.

Biofire LoD Genomic copies/ml Number of replicates Number of target detected
3/3 positive 2/3 positive 1/3 equivocal 0/3 negative
0.15 50 5 0 0 0 5
0.3 100 5 0 0 2 3
0.6 200 5 0 0 4 1
0.9 300 5 0 3 2 0
1.2 400 5 1 3 1 0
1.8 600 5 3 2 0 0
6 2,000 5 5 0 0 0