Table 3. Adequate protection rate, fit factor, and leakage rate between domestic and 3M-made N95 respirators.
Tested value | Manufacturing country | P value | OR | ||
---|---|---|---|---|---|
Made in Korea (n = 90) | Made by 3M (n = 60) | ||||
Fit factor | 14.6 | ||||
Pass, % (No.) | 25.6 (23) | 83.3 (50) | < 0.001 | ||
Fit factor, Median (IQR) | 38.5 (9.0–104.5) | 200.0 (152.0–200.0) | < 0.001 | ||
Leakage rate, % | 10.3 | ||||
Pass, % (No.) | 21.1 (19) | 73.3 (44) | < 0.001 | ||
Leakage rate, Median (IQR) | 8.9 (5.3–17.5) | 2.9 (1.1–5.1) | < 0.001 |