Skip to main content
. 2021 Jun 1;4:88. doi: 10.1038/s41746-021-00460-1

Table 4.

Diagnostic performance of non-gated model based on chosen ground truth method.

Cutoff score Non-gated routine chest CT Gated coronary CT
Sensitivity (%) Specificity (%) PPV (%) NPV (%) F1 Sensitivity (%) Specificity (%) PPV (%) NPV (%) F1
1 93.5 (86.4, 100) 100 (100, 100) 100 (100, 100) 88.2 (78.9, 97.5) 0.967 74.4 (61.7, 87.0) 100 (100, 100) 100 (100, 100) 41.2 (27.0, 55.4) 0.853
10 90.9 (82.6, 99.2) 100 (100, 100) 100 (100, 100) 92.3 (84.6, 100) 0.952 57.1 (42.8, 71.4) 100 (100, 100) 100 (100, 100) 42.3 (28.0, 56.6) 0.727
100 87.5 (77.9, 97.1) 97.4 (92.7, 100) 87.5 (77.9, 97.1) 97.4 (92.7, 100) 0.875 33.3 (19.7, 47.0) 96.0 (90.3, 100) 87.5 (77.9, 97.1) 63.2 (49.2, 77.1) 0.458
400 100 (100, 100) 100 (100, 100) 100 (100, 100) 100 (100, 100) 1.000 37.5 (23.5, 51.5) 100 (100, 100) 100 (100, 100) 88.4 (79.1, 97.6) 0.545

PPV positive predictive value, NPV negative predictive value.