Table 2.

Sensitivity and specificity compared to manufacturers' data from samples collected ≥14 days (Abbott, Roche) and >15 days (DiaSorin) post PCR confirmation.

  Study data
Manufacturers' dataa
  n Sensitivity % (95% CI) n Specificity % (95% CI) n Sensitivity % (95% CI) n Specificity % (95% CI)
Abbott 152 80.9 (74.0–86.4) 150 99.3 (96.3–100) 88 100 (95.9–100) 997 99.6 (99.0–99.9)
Roche 152 98.7 (95.3–99.8) 150 98.7 (95.3.99.8) 185 99.5 (97.0–100) 6305 99.8 (99.7–99.9)
DiaSorin 145 89.0 (82.8–93.1) 148 97.3 (93.3–98.9) 48 97.9 (89.1–99.6) 1200 98.6 (97.7–99.1)
Dynamiker 152 72.4 (64.8–78.9) 150 96.7 (92.4–98.6) 162 93.2 (not reported) 300 95.3 (not reported)

No information about the time of sampling was available for Dynamiker Biotechnology.

a

Data retrieved from following versions of product kit inserts: Abbott 6R86 H0791R04, June 2020; Roche V3, 2020-06; DiaSorin 200/007-797, 07, 2020-07; Dynamiker Biotechnology DNK-1419-1.